巜疯狂的少妇4做爰BD,成全视频在线观看在线播放高清,岳打开双腿开始配合交换,色999

Hotline
400-850-9596

Product current:home > Product > XJCJ-500M pipe drop hammer impact testing machine
XJCJ-500M pipe drop hammer impact testing machine

XJCJ-500M pipe drop hammer impact testing machine

First, overview of the instrument

It is suitable for the determination of impact toughness of PVC-U, PVC-C pipe, PP pipe, PE pipe, plastic pipe for underground communication pipe, etc. It is an ideal testing instrument for testing, production units, building materials industry and scientific research units.

Second, instrument characteristics

The testing machine has the function of preventing secondary impact, the use of AC servo system, lifting hammer speed, good positioning accuracy, high test efficiency, small air resistance, can automatically measure the height, drop hammer lifting height can be arbitrarily set, lifting hammer height automatic calibration, impact hammer weight can be assembled by weights, the capture rate of anti-secondary impact device is 100%. V-shaped iron adjustment to adapt to different pipe diameter of the sample, the hammer fall for free fall movement, according to the requirements of the order to provide to meet the other standards of the impact hammer, the hammer device, anti-secondary impact device using pneumatic transmission to achieve. Two test methods, pass method and gradient method, are applicable. High light (LED) digital tube display instant height; The lifting height of the drop hammer can be set arbitrarily, and the impact height can be set arbitrarily within the range of 50mm ~ 2000mm

(The capture rate of the anti-secondary impact device in this height range is 100%);

Automatic calibration of lifting height;

The impact weight can be assembled by weights;

The capture rate of the anti-secondary impact device is: 100%;

V-shaped iron adjustment to adapt to different diameters of pipes and different thickness of the plate samples, optional helmet accessories can be used for impact test helmet;

The fall of the hammer body is a free fall movement (friction-free introduction);

Impact hammers that meet other criteria are available on order;

The anti-secondary impact device adopts pneumatic capture;

The sample room is connected with the laboratory, and the temperature is the same as the test environment temperature;

Third, in line with standards

XJCJ-500M pipe drop hammer impact testing machine

JB/T9389-2008

Technical conditions for drop hammer impact testing machine for non-metallic materials

GB/T 14152-2001

External impact resistance of thermoplastic pipes

GB/T10002.1-2006

Hard polyvinyl chloride (PVC-U) pipe for water supply

GB/T13664-2006

Hard polyvinyl chloride (PVC-U) pipe for irrigation with low pressure water delivery

GB/T 16800-2008

Core layer foamed hard polyvinyl chloride (PVC-U) pipe for drainage

ISO4422, ISO3127, BSEN1411, BSEN744, etc.

Fourth, working principle

With the specified quality and size of the drop hammer from the specified height of the impact of the test sample specified parts, you can measure the batch (or continuous extrusion production) of the real impact rate of the product. This test method can be changed by changing the quality of the drop hammer or changing the height to meet the technical requirements of different products. The maximum allowable TIR value is 10%.

True impact rate:

When the whole batch of products are tested, the total number of impact damage divided by the total number of impacts is the real impact rate, expressed as a percentage.

Damage:

When observed with the naked eye, cracks, cracks or specimen breakage of the sample caused by impact are called damage, and the indentation or discoloration of the sample caused by the impact of the drop hammer is not considered damage. Host frame and guide rail:

Vertically fixed, can be adjusted and vertical, free release drop hammer, calibration, drop hammer impact pipe rate can not be less than 95% of the theoretical rate.

Hammer body:

The hammer body should comply with the provisions of Table 1 and Table 2, the hammer head should be made of steel, the minimum wall thickness is 5mm, the surface of the hammer head should not have dents, scratches and other visible defects affecting the test results, the weight of 0.5kg and 0.8kg drop hammer should have a d25 hammer, the mass of more than or equal to 1kg drop hammer should have a d90 hammer.

The specimen support:

Including a 120° Angle V-shaped bracket, its length should not be less than 200mm, and its fixed position should be such that the vertical projection of the impact point of the drop hammer should be within 2.5mm of the center line of the V-shaped bracket. When arbitrating the test, the lead screw top or support is used.

Release device:

The drop hammer can be dropped from any height at least 2 meters high, which means the height from the surface of the sample, ** to ±10mm. There should be a device to prevent the second impact of the drop hammer, and the rebound capture rate of the drop hammer should be guaranteed to 100%.

Table 1 Size of the hammer head of the drop hammer unit: mm

Model number

Rs

d±1

ds

a,(°)

d25

d90

50

50

25

90

Arbitrary

Any

 

Arbitrary

Any

 

Table 2 Recommended drop weight unit: kg

0.5

0.8

1.0

1.25

1.6

2.0

2.5

3.2

4.0

5.0

6.3

8.0

10.0

12.5

16.0

Note: The allowable tolerance for drop hammer mass is ±0.5%

Sample preparation:

The sample should be randomly cut from a batch or continuous production of pipes. The cutting end face should be perpendicular to the axis of the pipes. The cutting end should be clean and free of damage.

Sample length: The length of the sample is (200±10) mm

Sample marking:

Samples with an outside diameter greater than 40mm should be marked with equidistant lines along their length direction and numbered sequentially. The number of pipe samples with different outside diameters should be shown in Table 3. For pipes with an outside diameter less than or equal to 40mm, only one impact should be carried out on each sample. Table 3

Nominal outer diameter mm

Number of lines should be drawn

Nominal outside diameter mm

Number of lines should be drawn

40 or less

-

160

8

50

3

180

8

63

3

200

12

75

4

225

12

90

4

250

12

110

6

280

16

125

6

P 315

16

140

8

-

-

Condition conditioning:

The sample should be adjusted in a water bath or air bath at (0±1) ℃ or (20±2) ℃. The shortest adjustment time is shown in Table 4. Water bath should be used for arbitration test.

Table 4 State adjustment schedule of pipes with different wall thickness

Wall thickness δmm

Adjust time, min

Water bath

Air bath

The delta 8.6 or less

8.6 < delta 14.1 or less

The delta > 14.1

15

30

60

60

120

240

After state adjustment, the specimen with a wall thickness less than or equal to 8.6mm should be taken out of the air bath for 10s or from the water bath for 20s to complete the test. The specimen with wall thickness greater than 8.6mm should be taken out of the air bath within 20s or from the water bath within 30s to complete the test. If this interval is exceeded, the sample should be immediately put back into the pretreatment device and reprocessed for at least 5min. If the state adjustment temperature of the sample is (20±2) ℃ and the test environment temperature is (20±5) ℃, the time from taking out the sample to the completion of the test can be relaxed to 60s.

Fifth, test steps:

1, according to the provisions of product standards to determine the quality of the drop hammer and impact height.

2, the outer diameter of the sample is less than or equal to 40mm, each sample only withstand one impact.

3, the outer diameter of the sample is greater than 40mm in the impact test, first make the drop hammer impact on the No. 1 mark, if the sample is not damaged, then impact on the No. 2 mark, until the sample is damaged or all the marks are impacted once. Impact the sample one by one until the judgment result is obtained.

Sixth, technical parameters:

XJCJ-500M pipe drop hammer impact testing machine

Weight of hammer

0.250kg to 16.00kg in increments of 0.100kg and 0.250kg

Drop weight mass error

< 0.5% (where the accuracy of 2.000kg hammer and 1.000kg and 2.000kg weights is ±0.1%)

Diameter of hammer head

D90, D25; Note: (SR5, SR10, SR30) available according to user requirements.

Impact height

50mm~2000mm

Height indication error

< ±2mm

Specimen diameter

Φ10mm to Φ500mm

Lifting speed

10m/min

Hammer release control method

Pneumatic

Overall dimensions

Length 900mm x width 600mm x height 3410mm

Space required

Recommended front and back 1.5m, left and right 0.5m, upper 0.5m (minimum laboratory height is required to be 4m)

Seventh, instrument configuration

XJCJ-500M pipe drop hammer impact testing machine

Main engine

One

An electric box

One

A quick lift unit

A set

Hammer rod

Set (standard for execution provided by demander)

Press weights

A set (standard for execution provided by the demander)

Weights

A set (standard for execution provided by the demander)

Power cord

One

Medium V iron

A piece of

Small V-iron

One piece

Anchor screws

Four pieces

Wooden hands

One piece (standard according to demander)

 



the previous:Pipe static hydraulic testing machine next:XWW-20A electronic universal testing machine
Go top